FAQ / Question log

Technical questions as a log, not as a marketing accordion.

What is Kanzler Biologicals?

Kanzler Biologicals documents a commercial preclinical research and development project focused on artificial-heart engineering, mechanical circulatory support and regenerative cardiovascular research, active since 2017.

Is this a real, purchasable product?

No. This is preclinical engineering documentation — not an approved medical device and not medical advice. No treatment, purchase or clinical application is offered.

Why publish preclinical engineering details openly?

Transparency is the point: every value on this site is labelled as concept, target or confirmed, and open validation questions are listed rather than hidden.

How are the systemic and pulmonary circuits separated?

The systemic circuit needs higher outlet pressure and larger stroke volume; the pulmonary circuit runs at lower pressure with gentler flow. A single chamber for both tasks would either overload the lungs or under-supply the body.

What drives the systemic pump?

A pulsatile membrane pump (silicone-based membrane) via an electric linear drive (roller screw), serving the systemic circuit only: concept stroke 8–12 mm, chamber volume up to 130 ml, outlet pressure up to 210 mmHg (short-term design maximum, not a normal operating pressure).

What is the role of the pulse chamber?

It adds pulsatile energy to the continuous baseline flow and can act as a pressure buffer — it exists solely for pulsatility, not as a replacement for the main pump. Fill volume 40–50 ml; its beat frequency follows the main pump.

How does the microaxial pump work?

Approx. 21 mm diameter, approx. 10 mm body length, continuous operation with a rotor/impeller, pressure rise up to 20 mmHg, flow rate up to 13 l/min.

What is the bypass system for?

A two-stage safety relief: bypass valve 1 diverts flow toward the pulse chamber when outlet pressure or flow plausibility thresholds are exceeded; bypass valve 2 is a second safety path if bypass 1 is insufficient. The final relief route still needs architectural definition and hydraulic verification.

Why is pulmonary pressure kept so low?

Physiological pulmonary circulation runs well below systemic pressure; the microaxial pump is designed for a pressure rise up to 20 mmHg. The actual operating point must be patient-adaptive — people with pre-existing pulmonary hypertension run higher than any rigid limit.

What sensors monitor the system?

Pressure before/after the pump, flow, RPM, temperature, valve state and pulse chamber pressure — with cross-sensor plausibility checks (e.g. high RPM with low flow signals possible blockage, cavitation, clot or sensor drift).

What happens when a fault is detected?

A staged response: reduce power → bypass 1 → bypass 2 → alarm → defined safe mode. Each stage still requires bench and fault-injection testing.

Is the safety logic already validated?

No. Each stage of the staged response, and every sensor plausibility rule, still needs dedicated bench and fault-injection testing before it can be considered validated.

What powers the implant?

An implantable battery module: 14.8 V nominal, approx. 4.7 Ah, approx. 70 Wh (concept), with a BMS for cell balancing, thermal monitoring and overcharge/deep-discharge protection.

How is the battery recharged without a wired connection?

Through TET-based inductive charging: 3 equally sized external coils and 3 equally sized implanted receiver coils as 3 independent pairs (no intermediate coil), with 2 redundant electronics modules per unit. Frequency is still open (150 kHz is only a preliminary investigation value), target efficiency 85%, coil size still open, designed to minimise tissue heating.

What happens if charging is misaligned or interrupted?

Misalignment can turn electrical transfer into a thermal problem. Safe charge windows and stop criteria still need to be defined through tissue-equivalent phantom testing across distance, offset and angle.

What is confirmed versus still open?

Confirmed: pump dimensions, chamber volumes, battery voltage/capacity/energy, target efficiency. Still open, and explicitly labelled rather than guessed: TET frequency and coil size, heat-spreader fluid and pump, bus voltage (14.8 V vs. a possible ~22 V), emergency-pump electrical power, pulmonary operating pressure.

What would have to happen before any clinical claim?

Material validation, sterilisation strategy, biocompatibility, hemocompatibility, electrical safety, software validation, risk management and long-duration evidence — none of which exist yet.

Does this page make any medical claims?

No. Explicit disclaimer: preclinical R&D project, not an approved medical device, not medical advice. Evidence ratings and specs describe engineering targets, not clinical proof.